#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2621S (Expiry date : 02.05.2028)

Production date : 03.09.2026

Results of quality control 1 (after production), date: 09.09.2026

No serum blank (average) : 0.120

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.002 0.003 0.003 0.000 0
Negative control serum
(N° 9450-06)
0.056 0.054 0.065 0.062 0.059 3
Weak positive serum
(N° 9450-07)
0.785 0.805 0.799 0.812 0.800 35
Positive control serum
(N° 9450-08)
2.265 2.279 2.189 2.294 2.257 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.257 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 35 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.120 Yes

Comments:

Good for use.