#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2617D (Expiry date: 26.03.2028)
Production date : 27.07.2026
Results of quality control 1 (after
production), date: 29.07.2026
No serum blank (average) : 0.136
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
-0.004 |
0.009 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.103 |
0.095 |
0.105 |
0.084 |
0.096 |
4 |
Weak positive serum
(N° 9650-07) |
0.430 |
0.411 |
0.406 |
0.408 |
0.413 |
17 |
Positive control
serum
(N° 9650-08) |
2.291 |
2.472 |
2.347 |
2.422 |
2.383 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9650-08 |
> 1.200 |
2.383 |
Yes |
| Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
17 % |
Yes |
| Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.136 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Rue
de Morges 22,
CH-1023 Crissier, Switzerland.
Phone: + 41 21 633 31 67. Email: cb@bordier.ch
|