Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2616F (Expiry date: 21.03.2028)

Production date : 22.07.2026

Results of quality control 1 (after production), date: 24.07.2026

No serum blank (average) : 0.137

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.001 0.003 0.002 0.000 0
Negative control serum
(N° 9310-06)
0.028 0.022 0.032 0.029 0.027 1
Weak positive serum
(N° 9310-07)
0.152 0.161 0.142 0.136 0.147 7
Positive control serum
(N° 9310-08)
2.188 2.243 2.265 2.296 2.248 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.248 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 7 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.137 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Rue de Morges 22, CH-1023 Crissier, Switzerland.
Phone: + 41 21 633 31 67. Email: cb@bordier.ch