#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2611S (Expiry date : 03.01.2028)

Production date : 04.05.2026

Results of quality control 2 (after 6 months), date: 22.10.2026

No serum blank (average) : 0.139

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.003 0.003 -0.002 0.000 0
Negative control serum
(N° 9450-06)
0.055 0.052 0.045 0.051 0.050 2
Weak positive serum
(N° 9450-07)
0.677 0.713 0.618 0.698 0.676 31
Positive control serum
(N° 9450-08)
2.234 2.215 2.200 2.174 2.205 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.205 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 31 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 2 % Yes
Absorbance of no serum blank < 0.350 0.139 Yes

Comments:

Good for use.