Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2610C (Expiry date: 03.01.2028)

Production date : 04.05.2026

Results of quality control 2 (after 6 months), date: 22.10.2026

No serum blank (average) : 0.218

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 -0.001 0.007 -0.001 0.000 0
Negative control serum
(N° 9700-06)
0.167 0.171 0.168 0.169 0.169 7
Weak positive serum
(N° 9700-07)
0.455 0.462 0.431 0.493 0.461 20
Positive control serum
(N° 9700-08)
2.205 2.328 2.216 2.371 2.280 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.280 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 20 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 7 % Yes
Absorbance of no serum blank < 0.350 0.218 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch