#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2609B (Expiry date: 03.01.2028)

 

Production date : 04.05.2026

Results of quality control 2 (after 6 months), date: 22.10.2026

No serum blank (average) : 0.155

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.005 0.006 0.001 0.000 0
Negative control serum
(N° 9600-06)
0.064 0.063 0.059 0.061 0.062 3
Weak positive serum
(N° 9600-07)
0.486 0.454 0.485 0.495 0.480 21
Positive control serum
(N° 9600-08)
2.2167 2.308 2.225 2.304 2.251 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.251 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 21 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.155 Yes

Comments:

Good for use.