#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2602S (Expiry date : 15.10.2027)

Production date : 16.02.2026

Results of quality control 2 (after 6 months), date: 13.08.2026

No serum blank (average) : 0.141

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 -0.004 0.006 0.004 0.000 0
Negative control serum
(N° 9450-06)
0.071 0.084 0.107 0.094 0.089 3
Weak positive serum
(N° 9450-07)
0.973 0.938 1.008 0.963 0.970 38
Positive control serum
(N° 9450-08)
2.421 2.581 2.667 2.582 2.562 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.562 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 38 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.141 Yes

Comments:

Good for use.