#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2602S (Expiry date : 15.10.2027)
Production date : 16.02.2026
Results of quality control 1 (after
production), date: 19.02.2026
No serum blank (average) : 0.140
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.009 |
0.007 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.071 |
0.061 |
0.066 |
0.059 |
0.065 |
3 |
Weak positive serum
(N° 9450-07) |
0.862 |
0.836 |
0.846 |
0.853 |
0.850 |
36 |
Positive control serum
(N° 9450-08) |
2.202 |
2.509 |
2.419 |
2.409 |
2.385 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.385 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.140 |
Yes |
Comments:
Good for use.