Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2601T (Expiry date: 18.09.2027)

 

Production date : 19.01.2026

Results of quality control 1 (after production), date: 22.01.2026

No serum blank (average) : 0.127

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.005 0.003 0.004 0.000 0
Negative control serum
(N° 9200-06)
0.021 0.026 0.035 0.040 0.031 1
Weak positive serum
(N° 9200-07)
1.421 1.500 1.456 1.503 1.470 67
Positive control serum
(N° 9200-08)
2.148 2.202 2.169 2.310 2.208 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.208 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 67 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.127 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland