#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2539V (Expiry date: 02.08.2027)

 

Production date : 03.12.2025

Results of quality control 1 (after production), date: 11.12.2025

No serum blank (average) : 0.140

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.008 -0.006 0.006 0.008 0.000 0
Negative control serum
(N° 9400-06)
0.127 0.113 0.124 0.127 0.123 5
Weak positive serum
(N° 9400-07)
1.084 1.084 1.111 1.088 1.092 46
Positive control serum
(N° 9400-08)
2.260 2.446 2.295 2.421 2.356 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.356 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 46 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.140 Yes

Comments:

Good for use.