#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2537G (Expiry date: 26.07.2027)
Production date : 27.11.2025
Results of quality control 1 (after
production), date: 04.12.2025
No serum blank (average) : 0.126
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.003 |
-0.001 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.100 |
0.119 |
0.103 |
0.109 |
0.107 |
5 |
Weak positive serum
(N° 9350-07) |
1.133 |
1.099 |
1.196 |
1.169 |
1.149 |
55 |
Positive control serum
(N° 9350-08) |
2.045 |
2.115 |
1.998 |
2.125 |
2.070 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
2.070 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
55 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
5 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.