Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2536H (Expiry date: 26.07.2027)

 

Production date : 27.11.2025

Results of quality control 1 (after production), date: 03.12.2025

No serum blank (average) : 0.133

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.001 -0.003 0.007 0.000 0
Negative control serum
(N° 9550-06)
0.127 0.141 0.155 0.147 0.143 6
Weak positive serum
(N° 9550-07)
0.848 0.913 0.897 0.880 0.885 39
Positive control serum
(N° 9550-08)
2.185 2.328 2.245 2.283 2.260 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.260 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  17 % 39 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 6 % Yes
Absorbance of no serum blank < 0.350 0.133 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland