Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2533C (Expiry date: 18.07.2027)

Production date : 19.11.2025

Results of quality control 1 (after production), date: 25.11.2025

No serum blank (average) : 0.211

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.003 -0.001 0.003 0.000 0
Negative control serum
(N° 9700-06)
0.074 0.092 0.084 0.073 0.081 3
Weak positive serum
(N° 9700-07)
0.429 0.400 0.424 0.413 0.417 18
Positive control serum
(N° 9700-08)
2.266 2.362 2.322 2.361 2.328 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.328 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.211 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch