#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2531S (Expiry date : 15.06.2027)
Production date : 16.10.2025
Results of quality control 1 (after
production), date: 22.10.2025
No serum blank (average) : 0.134
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.003 |
-0.002 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.090 |
0.064 |
0.090 |
0.090 |
0.084 |
4 |
Weak positive serum
(N° 9450-07) |
0.864 |
0.854 |
0.855 |
0.828 |
0.851 |
36 |
Positive control serum
(N° 9450-08) |
2.270 |
2.351 |
2.291 |
2.419 |
2.333 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.333 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.