#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2529M (Expiry date: 31.05.2027)
Production date : 01.10.2025
Results of quality control 3 (after
12 months), date: 28.09.2026
No serum blank (average) : 0.136
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
0.001 |
0.004 |
-0.005 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.050 |
0.051 |
0.046 |
0.042 |
0.047 |
2 |
Weak positive serum
(N° 9300-07) |
0.840 |
0.864 |
0.820 |
0.765 |
0.822 |
34 |
Positive control serum
(N° 9300-08) |
2.391 |
2.613 |
2.399 |
2.401 |
2.451 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9300-08 |
> 1.200 |
2.451 |
Yes |
| Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
34 % |
Yes |
| Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.136 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|