Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2529M (Expiry date: 31.05.2027)

 

Production date : 01.10.2025

Results of quality control 3 (after 12 months), date: 28.09.2026

No serum blank (average) : 0.136

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.001 0.004 -0.005 0.000 0
Negative control serum
(N° 9300-06)
0.050 0.051 0.046 0.042 0.047 2
Weak positive serum
(N° 9300-07)
0.840 0.864 0.820 0.765 0.822 34
Positive control serum
(N° 9300-08)
2.391 2.613 2.399 2.401 2.451 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.451 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 34 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.136 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch