#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2529M (Expiry date: 31.05.2027)
Production date : 01.10.2025
Results of quality control 2 (after
6 months), date: 30.03.2026
No serum blank (average) : 0.135
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.006 |
0.007 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.047 |
0.059 |
0.055 |
0.055 |
0.054 |
2 |
Weak positive serum
(N° 9300-07) |
0.794 |
0.819 |
0.823 |
0.799 |
0.808 |
33 |
Positive control serum
(N° 9300-08) |
2.385 |
2.526 |
2.349 |
2.450 |
2.427 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9300-08 |
> 1.200 |
2.427 |
Yes |
| Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
33 % |
Yes |
| Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.135 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|