#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2529M (Expiry date: 31.05.2027)

 

Production date : 01.10.2025

Results of quality control 2 (after 6 months), date: 30.03.2026

No serum blank (average) : 0.135

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.006 0.007 0.001 0.000 0
Negative control serum
(N° 9300-06)
0.047 0.059 0.055 0.055 0.054 2
Weak positive serum
(N° 9300-07)
0.794 0.819 0.823 0.799 0.808 33
Positive control serum
(N° 9300-08)
2.385 2.526 2.349 2.450 2.427 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.427 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 33 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.135 Yes

Comments:

Good for use.