#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2529M (Expiry date: 31.05.2027)
Production date : 01.10.2025
Results of quality control 1 (after
production), date: 07.10.2025
No serum blank (average) : 0.131
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.001 |
-0.001 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.052 |
0.047 |
0.045 |
0.054 |
0.050 |
2 |
Weak positive serum
(N° 9300-07) |
0.738 |
0.713 |
0.775 |
0.743 |
0.742 |
32 |
Positive control serum
(N° 9300-08) |
2.246 |
2.365 |
2.398 |
2.294 |
2.326 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9300-08 |
> 1.200 |
2.326 |
Yes |
| Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
32 % |
Yes |
| Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|