#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2528B (Expiry date: 30.05.2027)

 

Production date : 01.10.2025

Results of quality control 1 (after production), date: 06.10.2025

No serum blank (average) : 0.145

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 0.003 0.002 -0.005 0.000 0
Negative control serum
(N° 9600-06)
0.071 0.060 0.060 0.067 0.064 3
Weak positive serum
(N° 9600-07)
0.510 0.505 0.488 0.459 0.490 23
Positive control serum
(N° 9600-08)
2.046 2.215 2.092 2.053 2.101 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.101 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 23 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.145 Yes

Comments:

Good for use.