#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2525A (Expiry date: 07.03.2027)

 

Production date : 08.07.2025

Results of quality control 3 (after 12 months), date: 07.07.2026

No serum blank (average) : 0.137

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.002 -0.007 0.009 0.000 0
Negative control serum
(N° 61
00-06)
0.098 0.101 0.094 0.082 0.094 4
Weak positive serum
(N° 61
00-07)
0.723 0.682 0.592 0.555 0.638 25
Positive control serum
(N° 61
00-08)
2.471 2.755 2.323 2.518 2.517 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.517 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 25 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.137 Yes

Comments:

Good for use.