#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2523V (Expiry date: 07.03.2027)

 

Production date : 08.07.2025

Results of quality control 3 (after 12 months), date: 07.07.2026

No serum blank (average) : 0.139

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 -0.001 0.002 0.003 0.000 0
Negative control serum
(N° 9400-06)
0.073 0.067 0.075 0.080 0.074 3
Weak positive serum
(N° 9400-07)
0.769 0.769 0.781 0.847 0.792 36
Positive control serum
(N° 9400-08)
2.160 2.212 2.189 2.312 2.219 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.219 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 36 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.139 Yes

Comments:

Good for use.