#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2517S (Expiry date : 03.02.2027)
Production date : 04.06.2025
Results of quality control 2 (after
6 months), date: 01.12.2025
No serum blank (average) : 0.139
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
0.000 |
0.000 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.089 |
0.088 |
0.078 |
0.074 |
0.082 |
4 |
Weak positive serum
(N° 9450-07) |
0.738 |
0.700 |
0.750 |
0.730 |
0.730 |
34 |
Positive control serum
(N° 9450-08) |
2.132 |
2.239 |
2.082 |
2.175 |
2.157 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.157 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
34 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.139 |
Yes |
Comments:
Good for use.