#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2515M (Expiry date: 15.12.2026)
Production date : 16.04.2025
Results of quality control 3 (after
12 months), date: 13.04.2026
No serum blank (average) : 0.152
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.002 |
0.003 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.013 |
0.024 |
0.013 |
0.022 |
0.018 |
1 |
Weak positive serum
(N° 9300-07) |
0.567 |
0.568 |
0.575 |
0.560 |
0.568 |
28 |
Positive control serum
(N° 9300-08) |
1.924 |
2.094 |
2.043 |
2.107 |
2.042 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9300-08 |
> 1.200 |
2.042 |
Yes |
| Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
28 % |
Yes |
| Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.152 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|