Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2515M (Expiry date: 15.12.2026)

 

Production date : 16.04.2025

Results of quality control 2 (after 6 months), date: 13.10.2025

No serum blank (average) : 0.143

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.007 0.001 0.005 0.000 0.000 0
Negative control serum
(N° 9300-06)
0.007 0.008 0.015 0.018 0.012 1
Weak positive serum
(N° 9300-07)
0.576 0.549 0.582 0.588 0.574 27
Positive control serum
(N° 9300-08)
2.065 2.166 2.116 2.149 2.124 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.124 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 27 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.143 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland