#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2514D (Expiry date: 15.12.2026)

 

Production date : 16.04.2025

Results of quality control 3 (after 12 months), date: 13.04.2026

No serum blank (average) : 0.141

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.002 -0.006 0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.088 0.082 0.077 0.055 0.076 4
Weak positive serum
(N° 965
0-07)
0.281 0.292 0.240 0.247 0.265 13
Positive control serum
(N° 9650
-08)
1.986 2.212 1.962 2.054 2.054 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.054 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 13 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.141 Yes

Comments:

Good for use.