#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2514D (Expiry date: 15.12.2026)
Production date : 16.04.2025
Results of quality control 2 (after
6 months), date: 14.10.2025
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.004 |
-0.001 |
0.008 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.064 |
0.067 |
0.057 |
0.072 |
0.065 |
3 |
Weak positive serum
(N° 9650-07) |
0.275 |
0.283 |
0.255 |
0.251 |
0.266 |
13 |
Positive control
serum
(N° 9650-08) |
1.912 |
2.107 |
1.909 |
1.944 |
1.968 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
1.968 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
13 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.