#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2513B (Expiry date: 26.11.2026)

 

Production date : 27.03.2025

Results of quality control 3 (after 12 months), date: 24.03.2026

No serum blank (average) : 0.154

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 -0.005 0.005 0.001 0.000 0
Negative control serum
(N° 9600-06)
0.061 0.059 0.045 0.066 0.058 3
Weak positive serum
(N° 9600-07)
0.366 0.391 0.374 0.381 0.378 18
Positive control serum
(N° 9600-08)
2.133 2.248 2.027 2.171 2.145 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.145 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 18 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.154 Yes

Comments:

Good for use.