#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2508S (Expiry date : 19.10.2026)
Production date : 20.02.2025
Results of quality control 3 (after
12 months), date: 19.02.2026
No serum blank (average) : 0.138
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.006 |
-0.003 |
0.007 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.059 |
0.065 |
0.078 |
0.080 |
0.071 |
3 |
Weak positive serum
(N° 9450-07) |
0.666 |
0.687 |
0.638 |
0.696 |
0.672 |
32 |
Positive control serum
(N° 9450-08) |
1.954 |
2.093 |
2.109 |
2.227 |
2.096 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.096 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.138 |
Yes |
Comments:
Good for use.