#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2506T (Expiry date: 21.09.2026)
Production date : 22.01.2025
Results of quality control 4 (after
20 months), date: 23.09.2026
No serum blank (average) : 0.130
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.007 |
-0.003 |
0.001 |
0.010 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.013 |
0.024 |
0.032 |
0.034 |
0.026 |
1 |
Weak positive serum
(N° 9200-07) |
1.259 |
1.078 |
1.177 |
1.185 |
1.175 |
57 |
Positive control serum
(N° 9200-08) |
1.974 |
2.098 |
2.028 |
2.120 |
2.055 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9200-08 |
> 1.200 |
2.055 |
Yes |
| Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
57 % |
Yes |
| Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.130 |
Yes |
Comments:
Good for use.