Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2506T (Expiry date: 21.09.2026)

 

Production date : 22.01.2025

Results of quality control 3 (after 12 months), date: 16.01.2026

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.002 0.001 0.001 0.000 0
Negative control serum
(N° 9200-06)
0.015 0.024 0.033 0.036 0.027 1
Weak positive serum
(N° 9200-07)
1.237 1.158 1.168 1.152 1.179 57
Positive control serum
(N° 9200-08)
2.005 2.041 2.065 2.144 2.064 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.064 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 57 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch