#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2504S (Expiry date : 14.09.2026)

Production date : 15.01.2025

Results of quality control 3 (after 12 months), date: 09.01.2026

No serum blank (average) : 0.128

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.001 -0.003 0.002 0.000 0
Negative control serum
(N° 9450-06)
0.078 0.079 0.060 0.075 0.073 4
Weak positive serum
(N° 9450-07)
0.652 0.652 0.627 0.653 0.646 32
Positive control serum
(N° 9450-08)
1.932 2.128 2.007 2.044 2.028 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.028 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 32 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.128 Yes

Comments:

Good for use.