#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2437S (Expiry date : 15.06.2026)
Production date : 16.10.2024
Results of quality control 3 (after
12 months), date: 10.10.2025
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.001 |
0.003 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.054 |
0.055 |
0.059 |
0.055 |
0.056 |
3 |
Weak positive serum
(N° 9450-07) |
0.659 |
0.631 |
0.671 |
0.656 |
0.654 |
32 |
Positive control serum
(N° 9450-08) |
2.011 |
2.151 |
2.053 |
2.050 |
2.066 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.066 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.