#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2430S (Expiry date : 03.05.2026)
Production date : 04.09.2024
Results of quality control 3 (after
12 months), date: 29.08.2025
No serum blank (average) : 0.130
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.002 |
0.000 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.062 |
0.085 |
0.078 |
0.081 |
0.077 |
4 |
Weak positive serum
(N° 9450-07) |
0.626 |
0.686 |
0.662 |
0.666 |
0.660 |
30 |
Positive control serum
(N° 9450-08) |
2.058 |
2.292 |
2.105 |
2.276 |
2.183 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.183 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
30 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.130 |
Yes |
Comments:
Good for use.