#9350 Echinococcus granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses

Lot 2428G (Expiry date: 13.04.2026)

 

Production date : 14.08.2024

Results of quality control 4 (after 20 months), date: 14.04.2026

No serum blank (average) : 0.134

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 0.001 0.001 -0.003 0.000 0
Negative control serum
(N° 9350-06)
0.034 0.040 0.042 0.033 0.037 2
Weak positive serum
(N° 9350-07)
0.866 0.789 0.878 0.886 0.854 42
Positive control serum
(N° 9350-08)
1.994 1.959 1.979 2.204 2.034 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9350-08 > 1.200 2.034 Yes
Absorbance of weak positive serum 9350-07, as % of positive serum >  27 % 42 % Yes
Absorbance of negative serum 9350-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.134 Yes

Comments:

Good for use.