#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2428G (Expiry date: 13.04.2026)
Production date : 14.08.2024
Results of quality control 4 (after
20 months), date: 14.04.2026
No serum blank (average) : 0.134
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
0.001 |
0.001 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.034 |
0.040 |
0.042 |
0.033 |
0.037 |
2 |
Weak positive serum
(N° 9350-07) |
0.866 |
0.789 |
0.878 |
0.886 |
0.854 |
42 |
Positive control serum
(N° 9350-08) |
1.994 |
1.959 |
1.979 |
2.204 |
2.034 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
2.034 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
42 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.