#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2422B (Expiry date: 25.02.2026)

 

Production date : 26.06.2024

Results of quality control 1 (after production), date: 01.07.2024

No serum blank (average) : 0.135

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.005 0.009 -0.001 0.000 0
Negative control serum
(N° 9600-06)
0.048 0.068 0.059 0.061 0.059 3
Weak positive serum
(N° 9600-07)
0.394 0.375 0.427 0.427 0.405 18
Positive control serum
(N° 9600-08)
2.138 2.328 2.194 2.325 2.246 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.246 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 18 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.135 Yes

Comments:

Good for use.