#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2411G (Expiry date: 28.10.2025)
Production date : 29.02.2024
Results of quality control 4 (after
20 months), date: 29.10.2025
No serum blank (average) : 0.116
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.002 |
0.004 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.025 |
0.023 |
0.035 |
0.034 |
0.029 |
2 |
Weak positive serum
(N° 9350-07) |
0.765 |
0.814 |
0.822 |
0.890 |
0.823 |
43 |
Positive control serum
(N° 9350-08) |
1.839 |
1.889 |
1.903 |
2.082 |
1.928 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
1.928 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
43 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.116 |
Yes |
Comments:
Good for use.