#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2322R (Expiry date: 12.03.2025)
Production date : 13.07.2023
Results of quality control 2 (after
6 months), date: 10.01.2024
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.004 |
-0.008 |
0.014 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.045 |
0.084 |
0.053 |
0.109 |
0.073 |
3 |
Weak positive serum
(N° 9750-07) |
0.452 |
0.471 |
0.434 |
0.480 |
0.460 |
21 |
Positive control serum
(N° 9750-08) |
2.108 |
2.230 |
2.123 |
2.201 |
2.166 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.166 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
21 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.