Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2322R (Expiry date: 12.03.2025)

 

Production date : 13.07.2023

Results of quality control 2 (after 6 months), date: 10.01.2024

No serum blank (average) : 0.126

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.004 -0.008 0.014 0.000 0
Negative control serum
(N° 9750-06)
0.045 0.084 0.053 0.109 0.073 3
Weak positive serum
(N° 9750-07)
0.452 0.471 0.434 0.480 0.460 21
Positive control serum
(N° 9750-08)
2.108 2.230 2.123 2.201 2.166 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.166 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 21 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.126 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch