Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2309S (Expiry date : 22.10.2024)

Production date : 23.02.2023

Results of quality control 2 (after 6 months), date: 22.08.2023

No serum blank (average) : 0.120

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.000 0.002 0.001 0.000 0
Negative control serum
(N° 9450-06)
0.086 0.087 0.090 0.094 0.089 4
Weak positive serum
(N° 9450-07)
0.770 0.753 0.788 0.778 0.772 36
Positive control serum
(N° 9450-08)
2.140 2.150 2.108 2.290 2.172 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.172 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 36 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.120 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bâtiment Biokema, CH-1023 Crissier, Switzerland