#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2309S (Expiry date : 22.10.2024)
Production date : 23.02.2023
Results of quality control 1 (after
production), date: 03.03.2023
No serum blank (average) : 0.121
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.007 |
0.005 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.110 |
0.104 |
0.116 |
0.109 |
0.110 |
5 |
Weak positive serum
(N° 9450-07) |
0.827 |
0.850 |
0.825 |
0.810 |
0.828 |
36 |
Positive control serum
(N° 9450-08) |
2.207 |
2.391 |
2.259 |
2.352 |
2.302 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.302 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.121 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|