#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2222G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus IgG ELISA, Article 9350
Expiry date: 11.03.2024, Control date: 10.01.2023,
Control # 2
No serum blank (average) : 0.112
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.004 |
0.004 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.029 |
0.053 |
0.052 |
0.053 |
0.046 |
2 |
Weak positive serum
(N° 9350-07) |
0.945 |
0.953 |
0.910 |
0.959 |
0.941 |
44 |
Positive control serum
(N° 9350-08) |
2.1.988 |
2.276 |
2.035 |
2.254 |
2.138 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.138 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.112 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|