#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2205G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus IgG ELISA, Article 9350
Expiry date: 06.10.2023, Control date: 05.08.2022,
Control # 2
No serum blank (average) : 0.123
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.003 |
0.006 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.031 |
0.041 |
0.042 |
0.049 |
0.041 |
2 |
Weak positive serum
(N° 9350-07) |
1.035 |
1.076 |
1.046 |
0.992 |
1.037 |
44 |
Positive control serum
(N° 9350-08) |
2.308 |
2.502 |
2.317 |
2.263 |
2.348 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.348 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.123 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|