#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for alveolar echinococcosis patient follow-up in humans
Lot 2202F
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus multilocularis (Em18)
IgG ELISA, Article 9310
Expiry date: 09.09.2023, Control date: 12.09.2023,
Control # 4
No serum blank (average) : 0.124
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
0.001 |
0.000 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
0.013 |
0.004 |
0.017 |
0.008 |
0.010 |
0 |
Weak positive serum
(N° 9310-07) |
0.078 |
0.093 |
0.093 |
0.094 |
0.089 |
4 |
Positive control serum
(N° 9310-08) |
2.076 |
2.216 |
2.098 |
2.182 |
2.143 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9310-08 |
> 1.200 |
2.143 |
Yes |
Absorbance of negative serum 9310-06, as % of positive serum |
< 5 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.124 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|