Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 2201M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis (Em2-Em18) IgG ELISA, Article 9300
Expiry date: 09.09.2023, Control date: 14.01.2022,  Control #  1

No serum blank (average) : 0.136

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.008 -0.002 0.003 0.006 0.000 0
Negative control serum
(N° 9300-06)
0.017 0.028 0.020 0.036 0.026 1
Weak positive serum
(N° 9300-07)
0.685 0.637 0.663 0.639 0.656 27
Positive control serum
(N° 9300-08)
2.381 2.404 2.410 2.487 2.421 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.421 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 12 % 1 % Yes
Absorbance of no serum blank < 0.350 0.136 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch