#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2129D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.07.2023, Control date: 11.07.2023,
Control # 4
No serum blank (average) : 0.117
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
0.001 |
0.001 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.071 |
0.060 |
0.059 |
0.054 |
0.061 |
3 |
Weak positive serum
(N° 9650-07) |
0.301 |
0.288 |
0.292 |
0.286 |
0.292 |
12 |
Positive control
serum
(N° 9650-08) |
2.308 |
2.433 |
2.425 |
2.525 |
2.423 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.423 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.117 |
Yes |
Comments:
Good for use.