#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2129D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.07.2023, Control date: 08.11.2022,
Control # 3
No serum blank (average) : 0.117
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.002 |
0.001 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.049 |
0.049 |
0.052 |
0.077 |
0.057 |
3 |
Weak positive serum
(N° 9650-07) |
0.233 |
0.221 |
0.242 |
0.245 |
0.235 |
11 |
Positive control
serum
(N° 9650-08) |
2.066 |
2.201 |
2.076 |
2.370 |
2.178 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.178 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.117 |
Yes |
Comments:
Good for use.