#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2117H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 16.02.2023, Control date: 13.06.2022,  Control #  3

No serum blank (average) : 0.145

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.006 0.003 0.003 0.000 0
Negative control serum
(N° 9550-06)
0.059 0.067 0.065 0.064 0.064 4
Weak positive serum
(N° 9550-07)
0.295 0.331 0.306 0.312 0.311 17
Positive control serum
(N° 9550-08)
1.715 1.836 1.811 1.800 1.791 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 1.791 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 4 % Yes
Absorbance of no serum blank < 0.350 0.145 Yes

Comments:

Good for use.