#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2116T
Results of quality control
Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 16.02.2023, Control date: 13.06.2022,
Control # 3
No serum blank (average) : 0.124
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.005 |
0.004 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.008 |
0.008 |
0.014 |
0.008 |
0.010 |
0 |
Weak positive serum
(N° 9200-07) |
1.138 |
1.158 |
1.053 |
1.099 |
1.112 |
56 |
Positive control serum
(N° 9200-08) |
1.959 |
2.061 |
1.956 |
2.009 |
1.997 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
1.997 |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 8 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.124 |
Yes |
Comments:
Good for use.