#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2116T
Results of quality control
Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 16.02.2023, Control date: 14.12.2021,
Control # 2
No serum blank (average) : 0.119
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.003 |
0.002 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.003 |
0.004 |
0.005 |
0.008 |
0.005 |
0 |
Weak positive serum
(N° 9200-07) |
1.162 |
1.167 |
1.223 |
1.239 |
1.197 |
56 |
Positive control serum
(N° 9200-08) |
2.067 |
2.136 |
2.117 |
2.170 |
2.122 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.122 |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 8 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.119 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|