#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2111G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus IgG ELISA, Article 9350
Expiry date: 05.01.2023, Control date: 09.01.2023,
Control # 4
No serum blank (average) : 0.113
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.005 |
-0.006 |
0.003 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.021 |
0.030 |
0.039 |
0.028 |
0.030 |
2 |
Weak positive serum
(N° 9350-07) |
0.800 |
0.825 |
0.813 |
0.833 |
0.818 |
46 |
Positive control serum
(N° 9350-08) |
1.673 |
1.897 |
1.662 |
1.858 |
1.773 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
1.773 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|