#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2108H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 10.11.2022, Control date: 03.09.2021,  Control #  2

No serum blank (average) : 0.146

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.012 0.002 -0.014 0.000 0.000 0
Negative control serum
(N° 9550-06)
0.095 0.085 0.088 0.081 0.087 5
Weak positive serum
(N° 9550-07)
0.392 0.384 0.372 0.391 0.385 20
Positive control serum
(N° 9550-08)
1.825 1.955 1.839 1.941 1.890 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 1.890 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 5 % Yes
Absorbance of no serum blank < 0.350 0.146 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch