#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2107D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.11.2022, Control date: 15.11.2022,
Control # 4
No serum blank (average) : 0.114
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.003 |
0.000 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.067 |
0.069 |
0.078 |
0.081 |
0.074 |
4 |
Weak positive serum
(N° 9650-07) |
0.196 |
0.224 |
0.212 |
0.221 |
0.214 |
12 |
Positive control
serum
(N° 9650-08) |
1.634 |
2.006 |
1.781 |
1.984 |
1.852 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
1.852 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.114 |
Yes |
Comments:
Good for use.